Quantification of RNA in HIV type 1 subtypes D and G by NucliSens and Amplicor assays in Abidjan, Ivory Coast

AIDS Res Hum Retroviruses. 1999 Apr 10;15(6):495-8. doi: 10.1089/088922299311015.

Abstract

We have compared the performance of the NucliSens and the standard and modified HIV Monitor assays to quantify HIV-1 RNA plasma viral load in 12 tuberculosis patients infected with HIV-1 env subtype D (n = 3) and env subtype G (n = 9) in Ivory Coast. RNA was quantified in all nine subtype G specimens by the modified Amplicor HIV Monitor (mean, 4.6 log10 copies/ml; range, 3.1-6.3 log10/ml), in seven specimens by NucliSens (mean, 4.4 log10 copies/ml; range, 2.7-5.5 log10 copies/ml), and in 6 specimens by the standard Amplicor HIV Monitor assay (mean, 4.2 log10 copies/ml; range, 3.5-5.0 log10 copies/ml). All three subtype D samples were amplified by both the modified Amplicor HIV Monitor (mean, 4.5 log10 copies/ml; range, 3.8-5.1 log10 copies/ml) and NucliSens (mean, 3.8 log10 copies/ml; range, 2.8-5.0 log10 copies/ml); two samples were quantified by the standard Amplicor HIV Monitor assay (mean, 3.0 log10 copies/ml; range, 2.4-3.6 log10 copies/ml). Our preliminary results suggest that the modified Amplicor HIV Monitor can accurately quantify HIV-1 RNA viral load in persons infected with subtype D and G strains.