Rapid analysis of parathion in biological samples using headspace solid-phase micro-extraction (HS-SPME) and gas chromatography/mass spectrometry (GC/MS)

Clin Chem Lab Med. 1999 Jun;37(6):639-42. doi: 10.1515/CCLM.1999.099.

Abstract

A simple and rapid method for the analysis of parathion in biological samples is presented. The method consists of the extraction of parathion from blood samples by headspace solid-phase micro-extraction (SPME), followed by capillary gas chromatography and mass spectrometry detection. The recoveries in the blood samples after addition of ammonium sulphate and sulphuric acid were between 85% and 89% compared to samples prepared in water. Linearity was established over a concentration range of 0.1-5 microg/g blood with acceptable coefficients of correlation and limits of detection reached 0.02-0.05 microg/g. The time for an analysis is 57 minutes for one sample, including the extraction step. In conclusion, HS-SPME in combination with GC/MS is an effective method for the determination and quantification of parathion-ethyl and parathion-methyl in biological material.

MeSH terms

  • Calibration
  • Cholinesterase Inhibitors / blood*
  • Gas Chromatography-Mass Spectrometry / methods
  • Hot Temperature
  • Humans
  • Insecticides / blood*
  • Parathion / blood*
  • Regression Analysis
  • Sensitivity and Specificity

Substances

  • Cholinesterase Inhibitors
  • Insecticides
  • Parathion