Determination of normal values using an automated coagulation timer for activated coagulation time and its application in dogs with hemophilia

J Vet Intern Med. 1999 Sep-Oct;13(5):433-6. doi: 10.1892/0891-6640(1999)013<0433:donvua>2.3.co;2.

Abstract

The present study was performed to determine normal values for the Medtronic HemoTec automated activated coagulation time (ACT) analyzer (Medtronic HemoTec Inc, Parker, CO, distributed in Switzerland by Convergenza AG, Vaduz, Liechtenstein), and to evaluate its ability to detect dogs with hemophilia. ACT was measured in 43 healthy dogs presented to the Companion Animal Hospital, University of Bern, Bern, Switzerland, with the Medtronic HemoTec ACT analyzer to determine normal values. The mean +/- 2 standard deviations (SDs) of the values obtained was defined as the normal range. ACT was measured 8-10 times on the same day in 6 dogs to determine repeatability. ACT also was measured in 11 dogs with hemophilia and compared with a conventional visual ACT measurement test and with the activated partial thromboplastin time (APTT). ACT values of the 43 dogs used to determine normal values ranged from 66.5 to 97.0 seconds (mean, 79.3 seconds; SD, 7.35 seconds; median, 78.5 seconds). A range of 64-95 seconds (mean +/- 2 SDs) was defined as the normal range for the tested device. Repeatability was poor (r = 0.256). ACT values measured with the automated device did not correlate with ACT values measured with a conventional visual test or with APTT Sensitivity of the test was 90.9%, specificity was 98.0%, and accuracy was 96.7%. Variability in the test results was large and may lead to incorrect results. The automated measurement device was not superior to the conventional visual method in evaluating dogs with hemophilia.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Dog Diseases / blood
  • Dog Diseases / diagnosis*
  • Dogs / blood*
  • Hemophilia A / blood
  • Hemophilia A / diagnosis
  • Hemophilia A / veterinary*
  • Partial Thromboplastin Time / veterinary
  • Reference Standards
  • Reference Values
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Whole Blood Coagulation Time / veterinary*