Six methods for the determination of C-peptide evaluated

Clin Chem Lab Med. 1999 Jul;37(7):745-51. doi: 10.1515/CCLM.1999.115.

Abstract

Six assays for C-peptide were evaluated. Five assays were radioimmunoassays; one was a chemiluminescence assay. The within-run coefficient of variation ranged between 2.1 and 22.9%. The methods correlated well with each other (r = 0.925 to 0.978). There were, however, proportional biases between the assays, in particular in the higher concentration range. All assays were linear up to at least 4.5 ng/ml C-peptide concentration. All assays were easy to perform. However, the systematic bias between the assays demonstrates that further effort is needed to standardize the measurement of C-peptide. In the mean-time, we recommend that laboratories conducting C-peptide measurements establish local, method-specific reference ranges.

Publication types

  • Comparative Study

MeSH terms

  • C-Peptide / blood*
  • Female
  • Humans
  • Immunoassay / methods*
  • Luminescent Measurements
  • Male
  • Radioimmunoassay

Substances

  • C-Peptide