MR-guided balloon angioplasty of stenosed aorta: in vivo evaluation using near-standard instruments and a passive tracking technique

J Magn Reson Imaging. 2000 Oct;12(4):639-44. doi: 10.1002/1522-2586(200010)12:4<639::aid-jmri18>3.0.co;2-w.

Abstract

The purpose of this study was to assess the feasibility of magnetic resonance (MR)-guided balloon angioplasty of a stenosed aorta on an open low-field magnet using a passive tracking technique. Visualization of vessels and position of instruments were realized by using a fast low-angle shot (FLASH) sequence. Catheters and guidewire were prepared for susceptibility-based MR visualization. Standard balloon catheters were inflated with diluted gadolinium, and nitinol guidewires were modified by incorporation of iron oxide markers into their walls. After validation on a flow phantom, balloon angioplasty was performed on an in vivo model of arterial stenosis. Creation of abdominal aorta stenosis was realized in five piglets. MR-guided balloon angioplasty of the aorta was performed with success in all but one. In one of them, stent implantation was achieved in the descending aorta. Balloon angioplasty using a passive tracking technique is a simple concept that can be realized with near-standard instruments and any MR imaging system. This represents an advance toward MR-guided vascular interventions in the future.

MeSH terms

  • Angioplasty, Balloon* / methods
  • Animals
  • Aorta, Abdominal
  • Aortic Diseases / therapy*
  • Constriction, Pathologic / therapy
  • Female
  • In Vitro Techniques
  • Magnetic Resonance Imaging*
  • Phantoms, Imaging
  • Stents
  • Swine