Variability of repeated coronary artery calcium measurements by electron beam tomography

Am J Cardiol. 2001 Jan 15;87(2):210-3, A8. doi: 10.1016/s0002-9149(00)01319-9.

Abstract

In 120 patients, the mean interscan variability of coronary calcium quantification by electron beam tomography was 19.9% (median 7.8%) for the traditional calcium score, and 16.2% (median 5.7%) for volumetric scoring. Although this difference was not significant, there was a significant influence of the total amount of calcium, number of acquired images, and image noise on interscan reproducibility.