The Cordis tantalum coil stent was assessed in a nonrandomized multicenter trial: 275 patients with stable or unstable angina were entered. Clinical follow-up was for 1 year, with repeat angiography at 6 months. The major adverse cardiac event rates (MACE) were 3%, 14%, and 17% at 1, 7, and 13 months, respectively. The procedural success rate was 96% and the subacute occlusion rate 1.5%, in a group of patients over 60% of whom had ACC/AHA type B2 or C lesions. The binary restenosis rate at 6 months was 17.3%. Minimum lumen diameter increased from 1.07 +/- 0.28 mm preprocedure to 2.93 +/- 0.34 mm poststenting and at 6 months was 1.99 +/- 0.69 mm. These results demonstrate that the Cordis tantalum stent can be used to treat complex lesions with good procedural success and low rates of subacute thrombosis and restenosis at 6 months.