Bipolar atrial sensing thresholds in sinus rhythm and atrial tachyarrhythmias. A comparative analysis in patients with DDDR pacemakers

Europace. 1999 Apr;1(2):135-9. doi: 10.1053/eupc.1998.0028.

Abstract

Automatic mode switching (AMS) function in dual chamber pacemakers depends on adequate detection of atrial tachyarrhythmias. There are few data on showing how intra-operative atrial signal amplititude during sinus rhythm can predict atrial tachyarrhythmias after pacemaker implantation. In 43 patients undergoing DDDR pacemaker implantation and atrioventricular nodal ablation for the treatment of drug-refractory paroxysmal atrial fibrillation, atrial sensing thresholds during sinus rhythm and during induced atrial tachyarrhythmias (24-48 h after device implantation) were analysed. Five different DDDR pacemaker systems were implanted (Chorus 7034, Ela Medical n = 13; Meta DDDR 1254, Telectronics Pacing Systems n = 12; Vigor DR 1230, Guidant n = 6; Trilogy DR 2364, Pacesetter, n = 2; Kappa DR 401, Medtronic USA n = 10). Every patient received a steroid-eluting, screwing, bipolar atrial lead (Medtronic, Capsure-Fix 4068). The mean P wave amplitude during implantation was 3.91 +/- 1.14 mV. The mean atrial sensing threshold during sinus rhythm and during all modes of induced atrial tachyarrhythmias was 3.35 +/- 1.0 mV, and 1.52 +/- 0.92 mV, respectively (P < 0.001). Atrial fibrillation was induced in 36 patients. The mean sensing threshold during sinus rhythm in this patient group was 3.39 +/- 1.01 mV, the mean sensing threshold during atrial fibrillation was 1.27 +/- 0.56 mV, reflecting a 63% reduction of sensing threshold compared with sinus rhythm (P < 0.001). Atrial flutter was induced in seven patients. The mean sensing threshold during sinus rhythm was 2.92 +/- 1.19 mV, the mean sensing threshold during atrial flutter was 2.79 +/- 1.26 mV, reflecting a reduction of 5% (ns) compared with sinus rhythm. Atrial sensing thresholds during sinus rhythm were significantly correlated with sensing thresholds during atrial tachyarrhythmias (r = 0.44; P < 0.002), but there were significant variations in intra-individual results. The reduction of atrial sensing thresholds between sinus rhythm and induced atrial tachyarrhythmias ranged from 30% to 82%.

Conclusion: Bipolar atrial sensing thresholds during sinus rhythm are correlated with sensing thresholds during atrial tachyarrhythmias, but there is a large degree of variance in individual patients. A 4:1 to 5:1 atrial sensing safety margin based on sensing threshold during sinus rhythm is a predictor for adequate postoperative detection of atrial tachyarrhythmias and the function of AMS devices.

Publication types

  • Comparative Study

MeSH terms

  • Arrhythmia, Sinus / physiopathology
  • Arrhythmia, Sinus / therapy
  • Electrophysiology
  • Heart / physiology
  • Humans
  • Middle Aged
  • Pacemaker, Artificial*
  • Stroke Volume / physiology
  • Tachycardia, Ectopic Atrial / therapy*