Fan-beam collimators offer a good balance between resolution and noise. The collimator response may be included in iterative reconstruction algorithms in order to improve single-photon emission tomography (SPET) resolution. To this end, accurate determination of the focal region and characterisation of the collimator response as a function of the source co-ordinates must be performed. In this paper, a method to characterise fanbeam collimators is evaluated. First, we calculated the real focal region and the accuracy of the collimator convergence. Then, we confirmed the hypothesis that Gaussian distributions adequately fit the collimator responses, although no individualised treatment was performed for the tails of detector response which are associated with scattering and septal penetration. Finally, analytical functions were used to model the resolution and sensitivity. The parameter values in these functions were obtained from experimental measures by non-linear regression fitting. Our findings show differences of 1.43% between nominal and real focal length and standard deviations of 2.5 mm in the x-direction and 7.1 mm in the y-direction for the focal convergence. The correlation coefficients between experimental and predicted values were 0.994 for resolution and 0.991 for sensitivity. As a consequence, the proposed method can be used to characterise the collimator response.