Second-generation Hybrid capture test and Amplicor monitor test generate highly correlated hepatitis B virus DNA levels

J Virol Methods. 2001 Sep;97(1-2):165-9. doi: 10.1016/s0166-0934(01)00359-7.

Abstract

The performance of the Digene Hybrid Capture II HBV DNA Test HC II and the Roche Cobas Amplicor Monitor Test (Cobas-HBV) was evaluated on 252 serum samples. One hundred and seventy-three samples were HBV DNA positive and 75 HBV DNA negative by both assays. Four samples were HBV DNA positive by Cobas-HBV only. Linear regression analysis showed that the HBV DNA concentrations obtained from both assays were significantly related (n=173, r=0.976, P<0.0001). The results of the study show that Hybrid capture II and Cobas-HBV could be used equally in the management for patients with chronic HBV infection.