Purpose: To examine the hypothesis that a transducer-tipped high-frequency ultrasound drug-delivery catheter may augment the thrombolytic effects of locally delivered low-dose urokinase and result in improved recanalization rates and reduced residual thrombotic burden.
Methods: Thrombi were induced in situ bilaterally in 5- to 6-cm-long segments of the superficial femoral arteries in 9 dogs by intraluminal thermal damage and injection of thrombin. A transducer-tipped high-frequency local drug-delivery catheter was applied at 1.1 MHz and 0.6 W for 60 minutes to one superficial femoral artery segment, and an identical catheter with an inactivated ultrasound transducer was used to treat the contralateral control segment. Urokinase (5000 IU/kg) was delivered bilaterally into the thrombi during the treatment interval.
Results: Angiography documented TIMI grade 2 or 3 flow in 9 (100%) segments in the ultrasound-treated group versus 6 (67%) of the controls (no ultrasound) (p = 0.058). Angiographically detected distal embolization was found in 2 ultrasound-treated segments compared with 5 controls (p = 0.02). Protruding or occlusive thrombi were seen angioscopically in 8 (89%) control segments but in only 1 (11%) of the ultrasound-treated arteries (p < 0.001). By histopathology, 7 (78%) segments in the control group had occlusive thrombi, whereas only 3 nonocclusive thrombi were found in the ultrasound-treatment group (p < 0.001).
Conclusions: Catheter-delivered high-frequency ultrasound and local low-dose urokinase infusion is efficacious for the treatment of acute thrombotic occlusions as evaluated by angiography, angioscopy, and histopathology.