Rapid evaluation of left ventricular volume and mass without breath-holding using real-time interactive cardiac magnetic resonance imaging system

J Am Coll Cardiol. 2001 Aug;38(2):527-33. doi: 10.1016/s0735-1097(01)01399-7.

Abstract

Objectives: The purpose of this study was to validate cardiac measurements derived from real-time cardiac magnetic resonance imaging (MRI) as compared with well-validated conventional cine MRI.

Background: Although cardiac MRI provides accurate assessment of left ventricular (LV) volume and mass, most techniques have been relatively slow and required electrocardiogram (ECG) gating over many heart beats. A newly developed real-time MRI system allows continuous real-time dynamic acquisition and display without cardiac gating or breath-holding.

Methods: Fourteen healthy volunteers and nine patients with heart failure underwent real-time and cine MRI in the standard short-axis orientation with a 1.5T MRI scanner. Nonbreath-holding cine MRI was performed with ECG gating and respiratory compensation. Left ventricular end-diastolic volume (LVEDV), left ventricular endsystolic volume (LVESV), ejection fraction (EF) and LV mass calculated from the images obtained by real-time MRI were compared to those obtained by cine MRI.

Results: The total study time including localization for real-time MRI was significantly shorter than cine MRI (8.6 +/- 2.3 vs. 24.7 +/- 3.5 min, p < 0.001). Both imaging techniques yielded good quality images allowing cardiac measurements. The measurements of LVEDV, LVESV, EF and LV mass obtained with real-time MRI showed close correlation with those obtained with cine MRI (LVEDV: r = 0.985, p < 0.001; LVESV: r = 0.994, p < 0.001; EF: r = 0.975, p < 0.001; LV mass: r = 0.977, p < 0.001).

Conclusions: Real-time MRI provides accurate measurements of LV volume and mass in a time-efficient manner with respect to image acquisition.

Publication types

  • Case Reports
  • Comparative Study
  • Research Support, Non-U.S. Gov't
  • Research Support, U.S. Gov't, P.H.S.
  • Validation Study

MeSH terms

  • Adult
  • Aged
  • Female
  • Heart Failure / diagnosis
  • Humans
  • Hypertrophy, Left Ventricular / diagnosis*
  • Image Interpretation, Computer-Assisted / methods
  • Magnetic Resonance Imaging / methods*
  • Magnetic Resonance Imaging, Cine
  • Male
  • Middle Aged
  • Observer Variation
  • Respiration
  • Stroke Volume
  • Time Factors
  • Ventricular Dysfunction, Left / diagnosis*