A strategy for the determination of enzyme kinetics using electrospray ionization with an ion trap mass spectrometer

Anal Chem. 2001 Nov 1;73(21):5078-82. doi: 10.1021/ac0105890.

Abstract

A simple and rapid means of enzyme kinetic analysis was achieved using electrospray ionization mass spectrometry and a one-point normalization factor. The model system used, glutathione S-transferase from porcine liver, is a two-substrate enzyme catalyzing the conjugation of glutathione with a variety of compounds containing an electrophilic center. An internal standard that is structurally similar to the product was added to the reaction quench solution, and a single-point normalization factor was used to determine the product concentration without the need of a calibration curve. Kinetic parameters, such as Km, Vmax and Ki (for thyroxine), obtained by electrospray mass spectrometry agreed with those obtained from traditional UV-vis spectroscopy, and competitive vs noncompetitive inhibition reactions could be delineated via mass spectrometry. These results suggest that our method can be applied to enzymatic processes in which spectrophotometric or spectrofluorometric assays are not feasible or when the relevant substrates do not incorporate chromophores or fluorophores. This new method is competitive with traditional UV assays in that it is facile and it involves very little analysis time.

Publication types

  • Validation Study

MeSH terms

  • Animals
  • Glutathione Transferase / antagonists & inhibitors
  • Glutathione Transferase / metabolism*
  • Kinetics
  • Liver / enzymology
  • Spectrometry, Mass, Electrospray Ionization / methods*
  • Swine

Substances

  • Glutathione Transferase