Candidate reference method for total thyroxine in human serum: use of isotope-dilution liquid chromatography-mass spectrometry with electrospray ionization

Clin Chem. 2002;48(4):637-42.

Abstract

Background: There is a need for a critically evaluated reference method for thyroxine to provide an accuracy base to which routine methods can be traceable. We describe a candidate reference method involving isotope-dilution coupled with liquid chromatography/mass spectrometry.

Methods: An isotopically labeled internal standard, thyroxine-d(5), was added to serum, followed by equilibration, protein precipitation, and ethyl acetate and solid-phase extractions to prepare samples for liquid chromatography-mass spectrometry electrospray ionization (LC/MS-ESI) analysis. For separation, a Zorbax Eclipse XDB-C(18) column was used with a mobile phase consisting of 1 mL/L acetic acid in acetonitrile-water (32:68 by volume) for positive ions and a Zorbax Extend-C(18) column with a mobile phase consisting of 2 mL/L ammonium hydroxide in methanol-water (32:68 by volume) for negative ions. [M + H](+) ions at m/z 778 and 783 for thyroxine and its labeled internal standard were monitored for positive ions and [M - H](-) ions at m/z 776 and 781 for negative ions. Samples of frozen serum pools were prepared and measured in three separate sets.

Results: Within-set CVs were 0.2-1.0%. The correlation coefficients of all linear regression lines (measured intensity ratios vs mass ratios) were 0.999-1.000. Positive- and negative-ion measurements agreed with a mean difference of 0.45% at three concentrations (50, 110, and 168 microg/L). The detection limits (at a signal-to-noise ratio of approximately 3 to 5) were 30 and 20 pg for positive and negative ions, respectively. The results from the LC/MS-ESI method were within 1 SD of the composite means from many routine clinical methods, although it appears that the clinical method means may be biased high by 4-5 microg/L across the concentrations. Some routine clinical methods may be biased by up to 20% at low concentrations.

Conclusions: This well-characterized LC/MS-ESI method for total serum thyroxine with a theoretically sound approach, demonstrated good accuracy and precision, and low susceptibility to interferences qualifies as a candidate reference method. Use of this reference method as an accuracy base may reduce the apparent biases in routine methods along with the high interlaboratory imprecision.

Publication types

  • Comparative Study

MeSH terms

  • Chromatography, Liquid / methods*
  • Chromatography, Liquid / statistics & numerical data
  • Humans
  • Linear Models
  • Quality Control
  • Reference Standards
  • Reference Values
  • Sensitivity and Specificity
  • Spectrometry, Mass, Electrospray Ionization / methods*
  • Spectrometry, Mass, Electrospray Ionization / statistics & numerical data
  • Thyroxine / blood*

Substances

  • Thyroxine