Background: A new intravenous contrast agent, AI-700, was evaluated to determine whether a bolus injection could be used to detect myocardial perfusion abnormalities during acute ischemia by using 2-dimensional (2D) and 3-dimensional (3D) myocardial contrast echocardiography.
Methods: 2D MCE was performed in 14 closed-chest dogs during coronary occlusion by using both continuous and triggered gray scale harmonic imaging and triggered power Doppler imaging. 3D MCE (open-chest) and nuclear perfusion imaging were performed in 10 of the 14 dogs. Postmortem triphenyl tetrazolium chloride (TTC) staining was performed to verify infarction.
Results: Thirteen of the 14 dogs had infarct by TTC; all 10 that had nuclear imaging showed a perfusion defect. Of the 13 dogs that had infarction, perfusion defects were detected in all (13 of 13) by gray scale harmonic imaging (sensitivity = 100%), and in 11 of 13 by power Doppler imaging (sensitivity = 85%). All 10 dogs that had nuclear imaging showed perfusion defects by gray scale harmonic imaging (sensitivity = 100%) and 8 of 10 by power Doppler imaging (sensitivity = 80%). The perfusion defect size, derived from 3D imaging (25% +/- 12%) correlated well with that from nuclear imaging (24% +/- 12%) (y = 0.9x + 3.8, r = 0.96, mean difference = 1.3% +/- 2.6%). The perfusion defect mass by 3D (22 +/- 14 g) also correlated well with the infarct mass by TTC staining (24 +/- 16 g) (y = 0.8x + 2.9, r = 0.89, P <.001, mean difference = -2.8 +/- 7.6 g).