[Determination of vanillymandelic acid in urine by high performance liquid chromatography]

Hunan Yi Ke Da Xue Xue Bao. 2000 Oct 28;25(5):492-4.
[Article in Chinese]

Abstract

Objective: To develop a high performance liquid chromatography(HPLC) for determining vanillylmandelic acid(VMA) in urine.

Methods: The analytical column was packed with Nova-Pak C18. Flow rate was 0.9 ml.min-1. The mobile phase was 40 mmol.L-1 NaH2PO4 buffer. The VMA of 30 healthy people in 24 hour urine were tested by the internal standard method and with electrochemical detector.

Result: The linear range was from 5 mumol.L-1 to 200 mumol.L-1. The mean analytical recovery rate was 99.1%. The mean within-day coefficient of variation(CV) was 5.8%, and the mean day-to-day CV was 7.7%. The detection limit was 0.28 mumol.L-1. The VMA of 24 hour urine was 18.13 +/- 6.17 mumol in healthy people.

Conclusions: The method is suitable for routine determination of VMA in urine.