BioMEMS device with integrated microdialysis probe and biosensor array

Biosens Bioelectron. 2002 Oct;17(10):859-65. doi: 10.1016/s0956-5663(02)00055-6.

Abstract

The fabrication of a microdevice for continuous sampling and on-line monitoring of glucose is described. The device comprised a microdialysis sampling system integrated on the flow through channel of a microfabricated enzyme sensor. The sensor was produced by thin film technology and was assembled to a printed circuit board (PCB) that provided the means for both electrical and fluidic connections. A polyacrilonitrile fibre, with a cut-off of 50 kDa, was used in the fabrication of the sampling probe. The performance of the device was evaluated in-vitro. High sampling efficiency of the microdialysis probe was achieved by appropriate selection of the perfusion fluid flow rate. Response times varying from 1.5 to 3.0 min were determined for flow rates ranging between 1 and 0.2 micro l/min. The linear response range was up to 30 mM glucose and interference from other electroactive substances was almost negligible. The device showed excellent stability under continuous operation for at least 5 days and sensitivity variation less than 3% over a period of 15 days.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Biosensing Techniques / instrumentation*
  • Calibration
  • Glucose / analysis*
  • Microdialysis / instrumentation*
  • Sensitivity and Specificity

Substances

  • Glucose