We compared eight HCV screening tests for detection of anti-HCV antibody; Ortho Quick Chaser HCV Ab (QC), Ortho HCV Ab ELISA III (ELISA), Ortho HVC Ab PA test III (PA), Lumipulse II Ortho HCV (LUMI), IMx HCV.DAINAPACKII (IMx), ARCHITECT HCV (ARCH), Immucheck.F-HCV C50 Ab (Immu), RANREAM HCV Ab Ex II (RAN). Sera from six hundred patients were examined by these eight screening tests. The positive rates of the eight screening tests were from 9.0% to 13.2%. Forty-five sera showed discrepant results between the eight screening tests, and about half of them showed weak positive reaction and/or false positive. Twenty-five of the forty-five sera were negative for ant-HCV antibody in the CHIRON RIBA III confirmatory test, and forty-four of them were negative for HCV-RNA in the PCR method. The agreement rates between the two reagents were from 95.5% to 99.2%, but were not always high between the two reagents that used similar antigen. The specificities and sensitivities evaluated by using the RIBA III confirmatory test were excellent in ELISA, LUMI, IMx, ARCH and Immu. Three BBI seroconversion panels were used to compare the positive readings in the initial stage of HCV infection by eight screening tests. ELISA and ARCH showed the earliest positive readings, and then IMx, LUMI = RAN, PA, QC and Immu in this order. These findings indicate that ELISA and ARCH were the most excellent in the sensitivity, specificity and early diagnosis of HCV infection. However, we must pay attention to the weak positive reaction in the screening tests, because there is a possibility of "false positive".