[Development of a gas chromatographic column system for the on-line analysis of trace tetrahydrofuran in hexane]

Se Pu. 2001 May;19(3):196-9.
[Article in Chinese]

Abstract

A gas chromatographic system using home-made 7 microns thick cross-linked dimethylpolysiloxane columns for the on-line analysis of trace tetrahydrofuran in hexane in a rubber production facility was developed. The experimental parameters of the column system, including the temperature, flow rate as well as the back-flushing and heart-cutting program were investigated. The total column-switching program was suggested. The system has been successfully operated for more than one year with good resolution, stability and precision (RSD < 5%) in analyzing trace tetrahydrofuran (0-250 x 10(-6), V/V) for industrial process control in the rubber production facility.

Publication types

  • English Abstract

MeSH terms

  • Chromatography, Gas / instrumentation*
  • Chromatography, Gas / methods
  • Dimethylpolysiloxanes
  • Furans / analysis*
  • Hexanes / chemistry

Substances

  • Dimethylpolysiloxanes
  • Furans
  • Hexanes