Electromagnetic actuator for generating variably oriented shear waves in MR elastography

Magn Reson Med. 2003 Jul;50(1):220-2. doi: 10.1002/mrm.10479.

Abstract

Magnetic resonance elastography (MRE) is a recently developed technique for determining the mechanical properties of biological tissue. In dynamic MRE, electromagnetic units (actuators) are widely used to generate shear waves in tissue. These actuators exploit the interaction between the static magnetic field B(0) and an annular coil supplied with alternating currents. Therefore, coil movements are restricted to selected orientations to B(0). Conventional actuators transfer this movement collinearly to B(0) into the tissue. In this study, an electromagnetic actuator was introduced that overcomes this limitation. It is demonstrated that different directions of mechanical excitation can be generated and monitored by MRE. Different spatial components of the propagation of the shear waves were determined using agarose phantoms. The technique allows maximum contrast for MRE images of objects with anisotropic strain components such as muscle tissue.

Publication types

  • Comparative Study
  • Evaluation Study
  • Research Support, Non-U.S. Gov't
  • Validation Study

MeSH terms

  • Anisotropy
  • Elasticity
  • Equipment Design
  • Equipment Failure Analysis
  • Magnetic Resonance Imaging / methods*
  • Magnetics / instrumentation*
  • Micromanipulation / instrumentation*
  • Micromanipulation / methods
  • Motion*
  • Muscle, Skeletal / physiology*
  • Physical Stimulation / instrumentation*
  • Physical Stimulation / methods*
  • Shear Strength
  • Vibration