[Study on the clinical use of alpha-fetoprotein (AFP) colloidal gold diagnostic strip]

Zhonghua Liu Xing Bing Xue Za Zhi. 2003 Mar;24(3):210-2.
[Article in Chinese]

Abstract

Objective: To test the sensitivity and specificity of alpha-fetoprotein (AFP) colloidal gold diagnostic strip as compared with RIA-AFP diagnostic kit.

Methods: Double blind method was used to test sera from 426 healthy people and 1567 patients selected in two general hospitals and a cancer hospital. Each serum was measured by RIA-AFP and AFP colloidal gold diagnostic strip on the same day. RIA-AFP diagnostic results was identified as true positive or negative.

Results: Among 426 healthy people, all the RIA-AFP test showed negative result (serum AFP concentration less than 25 IU/ml), but the AFP colloidal gold diagnostic strip had 1.88% false positive. When comparing the result from 1567 patients measured by RIA-AFP, the sensitivity and specificity of AFP colloidal gold diagnostic strip were 99.3%, and 97.2%, respectively. The crude correspondence rate between the two diagnostic regents was 97.6%.

Conclusion: AFP colloidal gold diagnostic strip showed very good result and could be used as a screening diagnostic kit in clinic and hospital settings.

Publication types

  • English Abstract
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Biomarkers, Tumor / blood*
  • Carcinoma, Hepatocellular / diagnosis*
  • Double-Blind Method
  • Gold Colloid
  • Humans
  • Liver Neoplasms / diagnosis*
  • Radioimmunoassay
  • Reagent Kits, Diagnostic*
  • Reproducibility of Results
  • Sensitivity and Specificity
  • alpha-Fetoproteins / analysis*

Substances

  • Biomarkers, Tumor
  • Gold Colloid
  • Reagent Kits, Diagnostic
  • alpha-Fetoproteins
  • Aurodye