Evaluation of the HiFocus electrode array with positioner in human temporal bones

J Laryngol Otol. 2003 Jul;117(7):527-31. doi: 10.1258/002221503322112932.

Abstract

The aim of the study presented was to assess the insertion mode and possible intracochlear trauma after implantation of the HiFocus electrode with positioner in human temporal bones. The study was performed in five freshly frozen temporal bones. The position of electrodes was evaluated using conventional X-ray analysis, rotational tomography and histomorphological analysis. Insertion of the HiFocus electrode with positioner resulted in considerable trauma to fine cochlear structures including fracture of the osseous spiral lamina, dislocation of the electrode array from the scala tympani into the scala vestibuli and fracture of the modiolus close to the cochleostomy. The implication of the results regarding clinical outcome will be discussed.

Publication types

  • Evaluation Study

MeSH terms

  • Cochlear Implantation / adverse effects
  • Cochlear Implantation / instrumentation*
  • Cochlear Implantation / methods
  • Electrodes, Implanted* / adverse effects
  • Fractures, Bone / etiology
  • Humans
  • Radiography
  • Spiral Lamina / injuries
  • Temporal Bone / diagnostic imaging
  • Temporal Bone / pathology
  • Temporal Bone / surgery*
  • Tomography / methods
  • Treatment Outcome