Experimental dosimetry of a 32P catheter-based endovascular brachytherapy source

Phys Med Biol. 2003 Aug 7;48(15):2283-96. doi: 10.1088/0031-9155/48/15/303.

Abstract

The experimental dosimetry in a water phantom of a 32P linear source, 20 mm in length, used for the brachytherapy of coronary vessels is reported. The source content activity, A, was determined by means of a calibrated well ion-chamber and the value was compared with the contained activity reported in the manufacturer's certification. In this field of brachytherapy dosimetry, radiochromic film supplies a high enough spatial resolution. A highly sensitive radiochromic film, that presents only one active layer, was used in this work for the source dosimetry in a water phantom. The radiochromic film was characterized by electron beams produced by a clinical linac. A Monte Carlo calculation of beta spectra in water at different distances along the source transverse bisector axis allowed to take into account the low dependence of film response from the electron beam energy. The adopted experimental set-up, with the source in its catheter positioned on the film plane inside the water phantom, supplies accurate dosimetric information. The measured dose rate to water per unit of source activity at reference distance, D(r0, theta0)/A, in units of cGy s(-1) GBq(-1), was in agreement with the value reported in the manufacturer's certification within the experimental uncertainty. The radial dose function, g(r), is in good agreement with the literature data. The anisotropy function F(r, theta) is also reported. The analysis of the dose profile obtained at 2 mm from the source longitudinal axis shows that the uniformity is within 10% along 75% of the 20 mm treatment length. The adopted experimental set-up seems to be adequate for the quality control procedure of the dose homogeneity distribution in the water medium.

Publication types

  • Comparative Study
  • Evaluation Study
  • Research Support, Non-U.S. Gov't
  • Validation Study

MeSH terms

  • Algorithms*
  • Brachytherapy / methods*
  • Calibration / standards
  • Catheters, Indwelling*
  • Coronary Restenosis / prevention & control
  • Coronary Restenosis / radiotherapy*
  • Europe
  • Film Dosimetry / instrumentation
  • Film Dosimetry / methods*
  • Film Dosimetry / standards*
  • Humans
  • Phosphorus Radioisotopes / analysis*
  • Radiometry / instrumentation
  • Radiometry / methods
  • Radiotherapy Dosage / standards
  • Radiotherapy Planning, Computer-Assisted / methods*
  • Reproducibility of Results
  • Sensitivity and Specificity

Substances

  • Phosphorus Radioisotopes