Comparison of isotope dilution mass spectrometry methods for the determination of total homocysteine in plasma and serum

Anal Chem. 2003 Sep 1;75(17):4631-8. doi: 10.1021/ac034207x.

Abstract

Two independent methods have been critically evaluated and applied to the measurement of total homocysteine in serum and plasma: solid-phase anion extraction (SPAE) gas chromatography/mass spectrometry (GC/MS) and protein precipitation liquid chromatography/tandem mass spectrometry (LC/MS/MS). In addition, analysis of samples prepared by SPAE was accomplished by liquid chromatography/mass spectrometry (LC/MS) and LC/MS/MS. These methods have been used to determine total homocysteine levels in several existing serum-based Standard Reference Materials (SRMs) from the National Institute of Standards and Technology and in patient plasma samples provided by the Centers for Disease Control and Prevention. The precision of the homocysteine measurements in serum and plasma was critically evaluated, and method comparisons were carried out using Bland-Altman plots and bias analysis. On the basis of the excellent precision and close agreement of the mass spectrometric (MS) methods, the MS-based methods will be used for certification of a serum-based SRM for homocysteine and folates.

Publication types

  • Research Support, U.S. Gov't, Non-P.H.S.

MeSH terms

  • Chromatography, Liquid
  • Gas Chromatography-Mass Spectrometry
  • Homocystine / blood*
  • Humans
  • Isotopes
  • Mass Spectrometry / methods*
  • Time Factors