Purpose: To investigate dose distributions in partial-volume irradiation experiments in small experimental animals, in particular the parotid gland of rat.
Materials and methods: High-resolution magnetic resonance imaging images were made that provided the outlines of the parotid glands, which were used to design collimators with conformal radiation ports for 100 and 50% cranial/caudal partial-volume irradiation. A protocol for absolute dosimetry was designed and relative dose measurements were performed. From the three-dimensional topographical data and the three-dimensional dose distribution, dose-volume histograms were determined.
Results: The standard uncertainty of absorbed entrance dose was within 3%. Radiochromic film, thermoluminescence dosemeters and ionization chamber dose measurements revealed that the relative doses measured were in good agreement. The 20-80% penumbra of the beam across the 50% field edge was only 0.4 mm at a 6 mm depth. The gradient of the percentage depth dose from the skin of the rat to a depth of 12 mm was 1.5% mm(-1). The absorbed doses in the cranial 50% and the caudal 50% partial volumes were comparable. This finding was reflected in the calculated dose-volume histograms of the different regions, which were similar. The dose in the shielded area between the left and right ports was about 14% of the dose near the centres of the beams.
Conclusion: The designed set-up showed that irradiation of small volumes could be performed with high accuracy allowing the study of differences in radiation damage. Similar doses were given to the 50% cranial and 50% caudal gland volumes and, therefore, a possible difference in radiosensitivity in these volumes was not a dose effect. The approach used was also applicable for the irradiation of small volumes of other tissues.