We propose an evaluation of a new heterogeneous immunoassay of cyclosporin on RXL HM Dimension (Dade Behring) for therapeutic cyclosporin monitoring in whole-blood patients transplant. The pretreatment step is performed automatically into the apparatus while it is a manual step with EMIT. Linearity, intra- and inter-day precision, limit of quantification, precision and accuracy of dilution steps and stability into the equipment were studied. We realized the comparison between ACMIA and EMIT methods on whole-blood patients transplant recipients. Heterogeneous immunoassay showed a good linearity between 0 and 500 ng/mL, intra- and inter-day precision with coefficient of variation inferior to 7.2%. We observed reproducible and accurate dilutions of high concentrations (500 to 2,000 ng/mL). The correlation with EMIT technique was correct for different type of transplant (n=55).