Simple method for the assay of linezolid in Brain Heart Infusion broth by high-performance liquid chromatography

Biomed Chromatogr. 2004 Jan;18(1):1-5. doi: 10.1002/bmc.283.

Abstract

A simple HPLC method was developed and validated for quantification of linezolid in Brain Heart Infusion (BHI) broth. Eperezolid was employed as internal standard and the sample pre-treatment procedure was simple. Calibration standards ranged from 0.05 to 16 mg/L. Accuracy was within 7.8% and reproducibility (RSD) was less than 8.3%. Recovery was approximately 100% at all concentrations examined. Linezolid was stable in the autosampler insert for at least 24 h at ambient temperature and in BHI for 72 h at 37 degrees C. This assay is rapid and ideal for analysis of a large number of samples.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Acetamides / analysis*
  • Anti-Infective Agents / analysis*
  • Chromatography, High Pressure Liquid / methods*
  • Culture Media / chemistry*
  • Linezolid
  • Oxazolidinones / analysis*
  • Reference Standards
  • Reproducibility of Results
  • Sensitivity and Specificity

Substances

  • Acetamides
  • Anti-Infective Agents
  • Culture Media
  • Oxazolidinones
  • Linezolid