Monte Carlo calculation of Varian 2300C/D Linac photon beam characteristics: a comparison between MCNP4C, GEANT3 and measurements

Appl Radiat Isot. 2005 Mar;62(3):469-77. doi: 10.1016/j.apradiso.2004.07.008.

Abstract

Different codes are used for Monte Carlo (MC) calculations in radiation therapy. In this research, MCNP4C and GEANT3 codes have been compared in calculations of dosimetric characteristics of Varian Clinac 2300C/D. The parameters of influence in the differences seen in dosimetric features were discussed. This study emphasizes that both MCNP4C and GEANT3 MC can be used in radiation therapy computations and their differences in photon spectra calculations have a negligible effect on percentage depth dose computations in radiation therapy.

Publication types

  • Comparative Study

MeSH terms

  • Algorithms
  • Humans
  • Monte Carlo Method*
  • Photons*
  • Radiation Dosage
  • Radiotherapy Planning, Computer-Assisted / methods*