Quantification of total and free mycophenolic acid in human plasma by liquid chromatography with fluorescence detection

J Chromatogr B Analyt Technol Biomed Life Sci. 2005 Mar 25;817(2):207-13. doi: 10.1016/j.jchromb.2004.12.005.

Abstract

A simple high-performance liquid chromatographic (HPLC) method was developed for the assay of total and free mycophenolic acid (MPA) in human plasma. Prior to analysis, total mycophenolic acid was extracted by protein precipitation and free drug was isolated from plasma samples using ultrafiltration. The extracts were injected onto a Kromasil C8 column at 30 degrees C with excitation and emission wavelengths set at 342 and 425 nm, respectively. The mobile phase was consisted of acetonitrile-32 mM glycine buffer, pH 9.2 (20:80, v/v), at a flow rate of 1.0 ml/min. The method was found to be linear over the concentration range investigated, 0.05-40 mg/l for total mycophenolic acid (r>0.999) and 5-1000 microg/l (r>0.99) for free drug. The percentage error of the analytical method was below 10.9%. The intra- and inter-day reproducibility was adequate with the coefficients of variation of 8.28% or below. The run time were 4 and 6 min for free and total MPA, respectively. The method thus can be effectively applied to measure mycophenolic acid concentrations in clinical samples.

Publication types

  • Research Support, Non-U.S. Gov't
  • Validation Study

MeSH terms

  • Chromatography, High Pressure Liquid / methods*
  • Humans
  • Hydrogen-Ion Concentration
  • Mycophenolic Acid / blood*
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Spectrometry, Fluorescence / methods*

Substances

  • Mycophenolic Acid