Preliminary evaluation of implantable MOSFET radiation dosimeters

Phys Med Biol. 2005 Jan 7;50(1):141-9. doi: 10.1088/0031-9155/50/1/011.

Abstract

In this paper, we report on measurements performed on a new prototype implantable radiation detector that uses metal-oxide semiconductor field effect transistors (MOSFETs) designed for in vivo dosimetry. The dosimeters, which are encapsulated in hermetically sealed glass cylinders, are used in an unbiased mode during irradiation, unlike other MOSFET detectors previously used in radiotherapy applications. They are powered by radio frequency telemetry for dose measurements, obviating the need for a power supply within each capsule. We have studied the dosimetric characteristics of these MOSFET detectors in vitro under irradiation from a 60Co source. The detectors show a dose reproducibility generally within 5% or better, with the main sources of error being temperature fluctuations occurring between the pre- and post-irradiation measurements as well as detector orientation. A better temperature-controlled environment leads to a reproducibility within 2%. Our preliminary in vitro results show clearly that true non-invasive in vivo dosimetry measurements are feasible and can be performed remotely using telemetric technology.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Calibration
  • Cobalt Radioisotopes
  • Dose-Response Relationship, Radiation
  • Equipment Design / methods*
  • Evaluation Studies as Topic
  • Film Dosimetry / methods
  • Humans
  • Phantoms, Imaging
  • Polystyrenes
  • Radio Waves
  • Radiometry / methods*
  • Radiotherapy Dosage
  • Reproducibility of Results
  • Semiconductors
  • Telemetry
  • Temperature
  • Time Factors
  • Tomography, X-Ray Computed

Substances

  • Cobalt Radioisotopes
  • Polystyrenes