Mechanical transient-based magnetic resonance elastography

Magn Reson Med. 2005 Mar;53(3):628-39. doi: 10.1002/mrm.20388.

Abstract

Magnetic resonance elastography (MRE) is a technique for quantifying material properties by measuring cyclic displacements of propagating shear waves. As an alternative to dynamic harmonic wave MRE or quasi-steady-state methods, the idea of using a transient impulse for mechanical excitation is introduced. Two processing methods to calculate shear stiffness from transient data were developed. The techniques were tested in phantom studies, and the transient results were found to be comparable to the harmonic wave results. Transient wave based analysis was applied to the brains of six healthy volunteers in order to assess the method in areas of complex wave patterns and geometry. The results demonstrated the feasibility of measuring brain stiffness in vivo using a transient mechanical excitation. Transient and harmonic methods both measure white matter (approximately 12 kPa) to be stiffer than gray matter ( approximately 8 kPa). There were some anatomic differences between harmonic and transient MRE, specifically where the transient results better depicted the deeper structures of the brain.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, U.S. Gov't, P.H.S.

MeSH terms

  • Algorithms
  • Brain Mapping*
  • Elastic Tissue / anatomy & histology*
  • Elasticity
  • Humans
  • Image Processing, Computer-Assisted
  • Magnetic Resonance Imaging / methods*
  • Phantoms, Imaging