Modelling simple helically delivered dose distributions

Phys Med Biol. 2005 Apr 7;50(7):1505-17. doi: 10.1088/0031-9155/50/7/013. Epub 2005 Mar 22.

Abstract

In a previous paper, we described quality assurance procedures for Hi-Art helical tomotherapy machines. Here, we develop further some ideas discussed briefly in that paper. Simple helically generated dose distributions are modelled, and relationships between these dose distributions and underlying characteristics of Hi-Art treatment systems are elucidated. In particular, we describe the dependence of dose levels along the central axis of a cylinder aligned coaxially with a Hi-Art machine on fan beam width, couch velocity and helical delivery lengths. The impact on these dose levels of angular variations in gantry speed or output per linear accelerator pulse is also explored.

Publication types

  • Evaluation Study

MeSH terms

  • Body Burden
  • Computer Simulation
  • Equipment Failure Analysis / methods
  • Models, Biological*
  • Models, Statistical*
  • Quality Assurance, Health Care / methods
  • Radiometry / methods*
  • Radiotherapy Dosage
  • Radiotherapy, Conformal / instrumentation*
  • Radiotherapy, Conformal / methods*
  • Relative Biological Effectiveness