Drug-eluting biodegradable poly-D/L-lactic acid vascular stents: an experimental pilot study

J Endovasc Ther. 2005 Jun;12(3):371-9. doi: 10.1583/05-1525.1.

Abstract

Purpose: To evaluate in vivo a new drug-eluting biodegradable vascular stent with respect to biocompatibility, neointimal hyperplasia formation, and reliability.

Methods: Self-expanding biodegradable poly-96L/4D-lactic acid (PLA) stents with 2 drugs (PLA + dexamethasone [DEX] and PLA + simvastatin [SIM]) and 2 different coatings (PLA + P(D,L)LA and PLA + polycaprolactone [PCL]) were compared with a self-expanding stainless steel Wallstent. The stents were implanted in both common iliac arteries of 8 pigs. Prior to sacrifice at 1 month, angiography was performed to determine patency. Specimens were harvested for quantitative histomorphometry; vascular injury and inflammation scores were assigned to the stented iliac segments.

Results: All stented arteries were angiographically patent. The mean luminal diameter (3.05 mm) and area (30.36 mm2) of DEX-eluting PLA stents were decreased compared to other stents (PLA + P(D,L)LA: 3.66 mm and 43.92 mm2; PLA + SIM: 4.21 mm and 56.48 mm2; PLA + PCL: 4.19 mm and 54.64 mm2; Wallstent: 5.01 mm and 81.19 mm2). Wallstents and DEX-eluting PLA stents induced minimal intimal hyperplasia: PLA + DEX: 0.16 mm, PLA + P(D,L)LA: 0.35 mm, PLA + SIM: 0.33 mm, PLA + PCL: 0.29 mm, and Wallstent: 0.18 mm. The vascular injury scores demonstrated only mild vascular trauma for all stents. Only mild to moderate inflammatory reaction was noted around stent struts with a vascular inflammation score.

Conclusions: Biodegradable polymer stents appear to be biocompatible and reliable, causing minimal neointimal hyperplasia. Furthermore, the new biodegradable poly-D/L-lactic acid stent can be used as a local drug delivery vehicle. The DEX-eluting PLA stent reduces neointimal hyperplasia. The findings show a need for further investigation to prove the efficacy and safety of this new biodegradable drug-eluting stent.

Publication types

  • Comparative Study
  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Absorbable Implants*
  • Animals
  • Arterial Occlusive Diseases / drug therapy
  • Arterial Occlusive Diseases / pathology
  • Arterial Occlusive Diseases / surgery
  • Blood Vessel Prosthesis Implantation / instrumentation*
  • Blood Vessel Prosthesis*
  • Coated Materials, Biocompatible*
  • Dexamethasone / pharmacology
  • Disease Models, Animal
  • Follow-Up Studies
  • Glucocorticoids / pharmacology
  • Hydroxymethylglutaryl-CoA Reductase Inhibitors / pharmacology
  • Iliac Artery / drug effects
  • Iliac Artery / pathology
  • Iliac Artery / surgery
  • Male
  • Pilot Projects
  • Polyesters*
  • Prosthesis Design
  • Simvastatin / pharmacology*
  • Stents*
  • Swine

Substances

  • Coated Materials, Biocompatible
  • Glucocorticoids
  • Hydroxymethylglutaryl-CoA Reductase Inhibitors
  • Polyesters
  • polycaprolactone
  • poly(lactide)
  • Dexamethasone