Evaluation of grey-scale standard display function as a calibration tool for diagnostic liquid crystal display monitors using psychophysical analysis

Med Biol Eng Comput. 2005 May;43(3):319-24. doi: 10.1007/BF02345807.

Abstract

The digital imaging and communications in medicine (DICOM) standard proposed the grey-scale standard display function (GSDF) as a calibration tool for making the gradation characteristic of a radiographic output image consistent. This is designed in such a manner that the contrast visually recognised by observers (called psychophysical contrast) becomes a constant for all digital driving levels. The DICOM standard calls such an ideal characteristic perceptual linearisation. The psychophysical gradient that can express the psychophysical contrast was introduced for the evaluation of the GSDF using a liquid crystal display monitor. Investigations regarding its ability to yield a constant psychophysical contrast, independent of the digital driving level change and under an actual observation environment, such as for clinical radiographic diagnosis in hospital, were carried out. The psychophysical gradients of the GSDF were obtained for two kinds of observation environment: one was a restricted environment such as in a dark room, under steady-state adaptation, using the sinusoidal grading pattern corresponding to the peak frequency of the human eye response. The other was an actual environment reflecting that encountered during clinical diagnosis in a hospital. As a result, the psychophysical gradient under the restricted environment became almost constant and independent of the change in digital driving level, i.e. perceptual linearisation could be satisfied. Furthermore, under the actual observation environment, the psychophysical gradient decreased gradually with the increase in digital driving level, i.e. the perceptual linearisation could not be satisfied. The percentage decrease in the value of the psychophysical gradient at the maximum luminance area was approximately 60% compared with that at the minimum luminance area. Accordingly, the GSDF is unsuitable as a calibration tool for the liquid crystal display monitor, which will be used in actual clinical diagnosis, as it cannot achieve 'perceptual linearisation' under the actual environment. For the purpose of clinical diagnosis, it is necessary to enlarge the physical gradient of GSDF further in the high digital driving level range (which relates to a high luminance area) to give an approximation that is as close to the idealised form as possible.

Publication types

  • Evaluation Study

MeSH terms

  • Calibration
  • Computer Terminals / standards*
  • Contrast Sensitivity
  • Data Display / standards*
  • Humans
  • Psychophysics
  • Radiographic Image Enhancement / standards*
  • Sensory Thresholds