Incidence of complications in patients with implantable cardioverter/defibrillator who receive additional transvenous pace/sense leads

Pacing Clin Electrophysiol. 2005 Aug;28(8):795-800. doi: 10.1111/j.1540-8159.2005.00169.x.

Abstract

Background: Implantation of an additional pace/sense (P/S) lead is commonly used in patients with implantable cardioverter/defibrillators (ICDs) to overcome P/S defects of integrated defibrillation leads (HV-P/S leads). No information is available about the clinical outcome and the incidence of complications in these patients.

Methods: Retrospective analysis was performed in 151 patients (125 male, age 54.9 +/- 13.6 years, LVEF 48.1 +/- 17.8%, CAD in 86 [57%], DCM in 24 [16%], ARVCM in 11 [7%]) who received an additional P/S lead between 1990 and 2002 (54 patients with abdominal and 97 patients with pectoral ICD system). Statistical analysis was done using Kaplan-Meier survival curves.

Results: The average follow-up (FU) after implantation of the additional P/S lead was 43 +/- 27 months. In total 117 patients [77.5%] remain implanted; 22 patients died due to cardiac-related reasons. After a FU of 23 +/- 23 months, 43 patients [28.5%] experienced lead-related problems after implantation of the additional P/S lead: oversensing in 23 [53.5%], insulation defect in 3 [7.0%], fracture in 1 [2.3%], system infection in 4 [9.3%], and defect of the HV-P/S lead in 6 [14.0%] patients. The event-free cumulative survival of the additional P/S lead after 1, 2, and 5 years was 87.0%, 79.8%, and 59.4%, respectively (for pectoral leads: 89.6%, 82.0%, and 60.0%, respectively).

Conclusions: Implantation of an additional P/S lead in case of failure of an HV-P/S lead is safe. However, it is associated with a substantial rate of complications during FU. Therefore, extraction of damaged defibrillation leads instead of implantation of P/S leads should be favored.

MeSH terms

  • Aged
  • Chi-Square Distribution
  • Defibrillators, Implantable*
  • Electrodes, Implanted*
  • Female
  • Humans
  • Incidence
  • Male
  • Middle Aged
  • Postoperative Complications / epidemiology*
  • Reoperation
  • Retrospective Studies
  • Statistics, Nonparametric