A new HPLC method for the determination of geniposide in rat serum with solid-phase extraction (SPE) for preconcentration is described. Geniposide and an internal standard (paeoniflorin) were extracted from serum by SPE using C18 cartridges. Analysis of the extract was then performed on a reversed-phase C18 column using acetonitrile-water (16:84, v/v) as the eluting solvent system, and UV detection at 238 nm was used to measure the analyte with a limit of quantitation about 0.1 microg/mL. The calibration curve for geniposide was linear (r = 0.9993) in the concentration range 0.1-16.0 microg/mL. The intra- and inter-day precision of the geniposide were determined and their RSD did not exceed 10%. The validated method has been successfully applied for pharmacokinetic studies of geniposide from rat serum after oral administration of Yin-Zhi-Ku decoction.
Copyright 2005 John Wiley & Sons, Ltd.