Thoracic vertebral screw impingement on the aorta in an in vivo bovine model

Spine (Phila Pa 1976). 2005 Nov 1;30(21):2406-13. doi: 10.1097/01.brs.0000184587.25330.40.

Abstract

Study design: A bovine model was used to evaluate the effects of thoracic vertebral screw impingement of the aorta.

Objectives: To evaluate the histologic and biomechanical changes in aortic wall tissue that was severely impinged by abutting instrumentation.

Summary of background data: Case reports of vascular injury associated with spinal instrumentation generally describe intraoperative injury; some report delayed presentation of large vessel damage. Risks associated with placing instrumentation adjacent to large vessels are largely unknown.

Methods: Six 1-month-old calves underwent left-sided thoracotomies, exposing the anterior thoracic spine and aorta. With the heads removed, screws were inserted in reverse fashion into T6 through T11, leaving the screw tips 1 cm proud and abutting the aorta. After 3, 6, or 12 months (2 calves each), the spines were resected with the adjacent aorta and underwent radiographic, histologic, and biomechanical testing.

Results: Computed tomography revealed varying degrees of vessel impingement. Although there were no frank ruptures, 96% of aortic specimens showed histopathologic changes, including 52% with wall thinning; 43% were no longer impinged, yet 60% of these had increased collagen (scar). Biomechanical testing of screw-impinged aortas demonstrated a lower failure stress (1.2 +/- 0.5 N/mm vs. 1.8 +/- 0.4 N/mm, P = 0.016) but no difference in failure strain (42 +/- 9% vs. 32 +/- 10%, P = 0.06) than controls.

Conclusions: Major impingement of vertebral screws on the aorta caused acute and chronic histopathologic and biomechanical changes in the vessel wall. This model represents a severe form of vessel penetration by a screw that confirms such a "worst case" scenario results in marked compromise of the vessel wall integrity. The sequelae of less severe impingement are unknown.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Aorta, Thoracic / injuries*
  • Aorta, Thoracic / pathology
  • Aorta, Thoracic / physiopathology
  • Bone Screws / adverse effects*
  • Cattle
  • Internal Fixators / adverse effects
  • Intraoperative Complications / etiology
  • Models, Animal
  • Spinal Fusion / adverse effects
  • Stress, Mechanical
  • Thoracic Vertebrae / diagnostic imaging
  • Thoracic Vertebrae / pathology
  • Thoracic Vertebrae / surgery*
  • Tomography, X-Ray Computed