High-performance liquid chromatographic method for determination of reversible isomers of SU5416

J Chromatogr A. 2006 Mar 31;1110(1-2):73-80. doi: 10.1016/j.chroma.2006.01.079. Epub 2006 Feb 10.

Abstract

SU5416 shows light-induced reversible geometric isomerism. A simple, reliable, isocratic HPLC method using an UV-vis detector at lambda(425nm) was developed. The method provides efficient (R(S)=3.5) analysis of the two isomers with retention of the isomeric integrity. Additionally, the method has linearity over a wide range (50-1000microg/mL, r(2)=0.99), is accurate (99-102%, RSD <4%), and reproducible (RSD <0.8%). The method was used for analyzing pharmaceutical samples and understanding the kinetics of SU5416 isomers in methanol. In addition, this method can be used for quantifying the non-isolatable E-isomer.

MeSH terms

  • Calibration
  • Chromatography, High Pressure Liquid / methods*
  • Indoles / analysis*
  • Indoles / chemistry
  • Isomerism
  • Kinetics
  • Pyrroles / analysis*
  • Pyrroles / chemistry
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Technology, Pharmaceutical / methods*