Reproducibility of intravascular ultrasound radiofrequency data analysis: implications for the design of longitudinal studies

Int J Cardiovasc Imaging. 2006 Oct;22(5):621-31. doi: 10.1007/s10554-006-9080-0. Epub 2006 Mar 31.

Abstract

Objectives: The purpose of this study was to assess in vivo the reproducibility of tissue characterization using spectral analysis of intravascular ultrasound (IVUS) radiofrequency data (IVUS-VH).

Background: Despite the need for reproducibility data to design longitudinal studies, such information remains unexplored.

Methods and results: IVUS-VH (Volcano Corp., Rancho Cordova, USA) was performed in patients referred for elective percutaneous intervention and in whom a non-intervened vessel was judged suitable for a safe IVUS interrogation. The IVUS catheters used were commercially available catheters (20 MHz, Volcano Corp., Rancho Cordova, USA). Following IVUS-VH acquisition, and after the disengagement and re-engagement of the guiding catheter, an additional acquisition was performed using a new IVUS catheter. Fifteen patients with 16 non-significant lesions were assessed by 2 independent observers. The relative inter-catheter differences regarding geometrical measurements were negligible for both observers. The inter-catheter relative difference in plaque cross-sectional area (CSA) was 3.2% for observer 1 and 0.5% for observer 2. The limits of agreement for (observer 1 measurements) lumen, vessel, plaque and plaque burden measurements were 0.82, -1.10 mm(2); 0.80, -0.66 mm(2); 1.08, -0.66 mm(2); and 5.83, -3.89%; respectively. Limits of agreement for calcium, fibrous, fibrolipidic and necrotic core CSA measurements were 0.22, -0.25 mm(2); 1.02, -0.71 mm(2); 0.61, -0.65 mm(2); and 0.43, -0.38 mm(2) respectively. Regarding the inter-observer agreement, the limits of agreement for lumen, vessel, plaque and plaque burden measurements were 2.61, -2.09 mm(2); 2.20-3.03 mm(2); 1.70, -3.04 mm(2); and 9.16, -16.41%; respectively, and for calcium, fibrous, fibrolipidic and necrotic core measurements of 0.08, -0.09 mm(2); 0.89, -1.28 mm(2); 0.74, -1.06 mm(2); and 0.16, -0.20 mm(2); respectively.

Conclusions: The present study demonstrates that the geometrical and compositional output of IVUS-VH is acceptably reproducible.

Publication types

  • Comparative Study

MeSH terms

  • Coronary Disease / diagnostic imaging*
  • Coronary Vessels / diagnostic imaging*
  • Female
  • Humans
  • Image Interpretation, Computer-Assisted* / methods
  • Longitudinal Studies
  • Male
  • Middle Aged
  • Observer Variation
  • Prospective Studies
  • Reproducibility of Results
  • Research Design
  • Ultrasonography, Interventional / instrumentation*