Transient ST-segment-elevation during pulmonary vein ablation using circumferential coiled microelectrodes in a prospective multi-centre study

Europace. 2006 Mar;8(3):178-81. doi: 10.1093/europace/euj013. Epub 2006 Jan 5.

Abstract

Aims: Paroxysmal atrial fibrillation (PAF) is predominantly triggered by focal ectopies located within the pulmonary veins (PV). The BITMAP Study (Breakthrough and Isolation Trial: Mapping and Ablation of Pulmonary Veins) investigated prospectively the safety and efficacy of a catheter design with circumferential mapping and ablation electrodes. We report the phenomenon of ST-segment-elevation during catheter placement in the left atrium (LA) and superior PVs in this multi-centre study.

Methods and results: Forty-three patients (57+/-10 years) with PAF were included in this study. Radiofrequency catheter (RFC) ablation supported by the 4F REVELATION Helix microcatheter (Cardima Inc., Freemont, CA, USA) with eight distal-coiled microelectrodes for bipolar mapping and ablation. RFC was applied at the ostial region of PV (30 W, 45-50 degrees C) with a maximum of four RFC applications per electrode. In four of the 43 patients from three centres, we recorded the occurrence of ST-segment-elevation greater than 0.2 mV and accompanying left thoracic discomfort. The ECG changes and the symptoms started abruptly and lasted for 4.2+/-2.2 min. Pericardial effusion could instantaneously be excluded by echocardiography in all cases. Coronary angiograms were performed in three patients with the longest episodes; no thrombotic material or air emboli were present. The symptoms and the ECG changes resolved completely in all patients.

Conclusion: The phenomenon of ST-segment-elevation during LA- and PV-mapping in patients with PAF may be a common occurrence. In this prospective multi-centre trial, we demonstrated the reversibility of this phenomenon; no cardiovascular or cerebral damage was reported during both the procedure and the follow-up. Although the mechanism is still unclear, vasospasm may contribute to this phenomenon because of autonomic dysregulation.

Publication types

  • Multicenter Study

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Atrial Fibrillation / surgery*
  • Cardiac Catheterization
  • Catheter Ablation / adverse effects*
  • Catheter Ablation / instrumentation
  • Electrocardiography*
  • Female
  • Heart Conduction System / surgery*
  • Humans
  • Male
  • Microelectrodes*
  • Middle Aged
  • Pulmonary Veins*