Comparative evaluation of the performance of the Abbott real-time human immunodeficiency virus type 1 (HIV-1) assay for measurement of HIV-1 plasma viral load following automated specimen preparation

J Clin Microbiol. 2006 May;44(5):1788-91. doi: 10.1128/JCM.44.5.1788-1791.2006.

Abstract

The performance of the new Abbott real-time human immunodeficiency virus type 1 (HIV-1) assay for HIV-1 RNA load determination in plasma was compared to that of the Abbott LCx HIV-1 RNA quantitative assay following automated RNA isolation by the Abbott m1000 extractor. The measured viral loads of 89 clinical specimens differed by mean 0.19 log10 copies/ml (95% confidence interval, 0.12 to 0.26 log10 copies/ml). Although the difference in viral load determinations was positively skewed in favor of the LCx assay, it did not reach statistical significance (P = 0.42). Results were linearly associated (R2 = 0.94) and strongly correlated (R = 0.96). Good performance was observed with HIV-1 subtypes other than B and circulating recombinant forms, although results obtained with two subtype G specimens and one H specimen showed a more substantial difference.

Publication types

  • Comparative Study
  • Evaluation Study

MeSH terms

  • Automation
  • HIV Infections / virology
  • HIV-1 / classification
  • HIV-1 / genetics
  • HIV-1 / isolation & purification*
  • Humans
  • RNA, Viral / blood
  • RNA, Viral / genetics
  • Reproducibility of Results
  • Viremia / virology
  • Virology / methods*
  • Virology / statistics & numerical data

Substances

  • RNA, Viral