Alumina ceramic-on-ceramic total hip replacement with a layered acetabular component

J Bone Joint Surg Br. 2006 Jul;88(7):877-82. doi: 10.1302/0301-620X.88B7.17675.

Abstract

A modular layered acetabular component (metal-polyethylene-ceramic) was developed in Japan for use in alumina ceramic-on-ceramic total hip replacement. Between May 1999 and July 2000, we performed 35 alumina ceramic-on-ceramic total hip replacements in 30 consecutive patients, using this layered component and evaluated the clinical and radiological results over a mean follow-up of 5.8 years (5 to 6.5). A total of six hips underwent revision, one for infection, two for dislocation with loosening of the acetabular component, two for alumina liner fractures and one for component dissociation with pelvic osteolysis. There were no fractures of the ceramic heads, and no loosening of the femoral or acetabular component in the unrevised hips was seen at final follow-up. Osteolysis was not observed in any of the unrevised hips. The survivorship analysis at six years after surgery was 83%. The layered acetabular component in our experience, has poor durability because of unexpected mechanical failures including alumina liner fracture and component dissociation.

MeSH terms

  • Acetabulum / diagnostic imaging
  • Acetabulum / surgery*
  • Aged
  • Aged, 80 and over
  • Aluminum Oxide*
  • Arthritis / diagnostic imaging
  • Arthritis / surgery*
  • Arthroplasty, Replacement, Hip / instrumentation*
  • Ceramics*
  • Female
  • Hip Joint / diagnostic imaging
  • Hip Joint / surgery
  • Humans
  • Middle Aged
  • Osteoarthritis, Hip / diagnostic imaging
  • Osteoarthritis, Hip / surgery
  • Postoperative Complications
  • Prosthesis Design
  • Prosthesis Failure
  • Radiography
  • Reoperation
  • Survival Analysis
  • Treatment Outcome

Substances

  • Aluminum Oxide