Comparison of BMD precision for Prodigy and Delphi spine and femur scans

Osteoporos Int. 2006;17(9):1303-8. doi: 10.1007/s00198-006-0127-9. Epub 2006 Jul 6.

Abstract

Introduction: Precision error in bone mineral density (BMD) measurement can be affected by patient positioning, variations in scan analysis, automation of software, and both short- and long-term fluctuations of the densitometry equipment. Minimization and characterization of these errors is essential for reliable assessment of BMD change over time.

Methods: We compared the short-term precision error of two dual-energy X-ray absorptiometry (DXA) devices: the Lunar Prodigy (GE Healthcare) and the Delphi (Hologic). Both are fan-beam DXA devices predominantly used to measure BMD of the spine and proximal femur. In this study, 87 women (mean age 61.6+/-8.9 years) were measured in duplicate, with repositioning, on both systems, at one of three clinical centers. The technologists were International Society for Clinical Densitometry (ISCD) certified and followed manufacturer-recommended procedures. All scans were acquired using 30-s scan modes. Precision error was calculated as the root-mean-square standard deviation (RMS-SD) and coefficient of variation (RMS-%CV) for the repeated measurements. Right and left femora were evaluated individually and as a combined dual femur precision. Precision error of Prodigy and Delphi measurements at each measurement region was compared using an F test to determine significance of any observed differences.

Results: While precision errors for both systems were low, Prodigy precision errors were significantly lower than Delphi at L1-L4 spine (1.0% vs 1.2%), total femur (0.9% vs 1.3%), femoral neck (1.5% vs 1.9%), and dual total femur (0.6% vs 0.9%). Dual femur modes decreased precision errors by approximately 25% compared with single femur results.

Conclusions: This study suggests that short-term BMD precision errors are skeletal-site and manufacturer specific. In clinical practice, precision should be considered when determining: (a) the minimum time interval between baseline and follow-up scans and (b) whether a statistically significant change in the patient's BMD has occurred.

Publication types

  • Comparative Study
  • Multicenter Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Absorptiometry, Photon / instrumentation*
  • Aged
  • Aged, 80 and over
  • Bone Density*
  • Female
  • Femur / physiology*
  • Femur Neck / physiology
  • Hip Joint / physiology
  • Humans
  • Lumbar Vertebrae / physiology*
  • Middle Aged
  • Osteoporosis, Postmenopausal / diagnosis
  • Postmenopause / physiology
  • Reproducibility of Results