Quantification of HIV-1 RNA load by one-tube-one-step RT PCR and real time PCR assay with TaqMan probe

J Med Assoc Thai. 2005 Nov:88 Suppl 3:S206-13.

Abstract

Objectives: To develop a less expensive assay to calculate HIV-1 viral load for use in resource-limited countries.

Material and method: An In-house One-tube-one-step Viral Load Assay (IOVA) was developed by using real-time PCR-based with TaqMan probe. Primers and probe were designed from the conserved region of sequences from all HIV subtypes. A standard curve was generated from reference virus in various dilutions. IOVA was applied on 105 HIV-positive and 25 HIV-negative samples and compared with the results from ROCHE AMPLICLOR.

Results: IOVA measured HIV RNA in the samples ranging from 125 to 2 x 10(6) copies/mL. The coefficient of variation of intra- and inter-assay ranged from 0.68% to 7.89%. The sensitivity, specificity, positive and negative predictive values were 92%, 100%, 100% and 79.5% respectively. The parallel quantitative analysis showed high correlation (r=0.95) between IOVA and AMPLICOR.

Conclusion: A new HIV-1 viral load assay was developed and validated. It was reliable and less expensive.

MeSH terms

  • Adult
  • Child
  • HIV Infections / blood*
  • HIV-1 / genetics*
  • Humans
  • RNA Probes
  • RNA, Viral / blood*
  • Reproducibility of Results
  • Reverse Transcriptase Polymerase Chain Reaction
  • Sensitivity and Specificity
  • Viral Load / methods*