There are a variety of arthroscopic devices used to pass sutures through the rotator cuff for its repair. Because they vary in size and shape, it is possible that they could damage the cuff and affect the integrity of the repair. We chose four devices for assessment--SutureLasso (Arthrex, Naples, FL), straight BirdBeak (Arthrex, Naples, FL), Viper (Arthrex, Naples, FL), and a #7 tapered Mayo needle--and performed cuff reattachments in four paired shoulders using suture anchors. These repairs were cycled and tested to failure. The SutureLasso and Mayo needle repairs failed at approximately 285 N whereas the BirdBeak and Viper failed during cycling at 150 N. It appears that the devices, which made the bigger holes in the cuff, can compromise the integrity of the repair.