Evaluation of neuromuscular transmission by using monopolar needle electrode

Acta Neurol Scand. 2006 Nov;114(5):340-5. doi: 10.1111/j.1600-0404.2006.00634.x.

Abstract

Objective: To evaluate the value of single-fibre electromyography (SFEMG) with monopolar electrode (MNPE) in revealing neuromuscular transmission dysfunction.

Material and methods: We examined the extensor digitorum communis muscle by using single-fibre electrode (SFE) and MNPE sequentially, in randomly assigned 20 healthy volunteers and in 17 patients with known myasthenia gravis (MG). The high-pass filter setting was 3 kHz for MNPE. Ten individual jitter values were calculated for each electrode in every muscle. Repetitive nerve stimulation (RNS) test on trapezius muscle was performed on 15 patients.

Results: In controls, the mean jitter values were 27 +/- 9 (10-59) micro s with SFE, and 21 +/- 7.2 (9-56) micro s with MNPE (P = 0.001). In the MG group, the mean jitter values were 52.4 +/- 38 (12-221) micro s with SFE, and 51.8 +/- 34.7 (12-179) micro s with MNPE. Both electrodes identified junction dysfunction in 14 patients. RNS revealed decrement in four patients but 11.

Conclusion: SFEMG with SFE is still the gold standard; however, SFEMG with MNPE is superior to RNS like SFEMG with SFE.

Publication types

  • Comparative Study
  • Evaluation Study

MeSH terms

  • Adolescent
  • Adult
  • Aged
  • Electric Stimulation
  • Electrodes / standards
  • Electromyography / instrumentation*
  • Electromyography / methods
  • Female
  • Humans
  • Male
  • Middle Aged
  • Muscle Contraction / physiology
  • Muscle, Skeletal / innervation
  • Muscle, Skeletal / physiopathology
  • Myasthenia Gravis / diagnosis*
  • Myasthenia Gravis / physiopathology*
  • Neuromuscular Junction / physiology
  • Neuromuscular Junction / physiopathology*
  • Predictive Value of Tests
  • Reference Values