Background: The number of patients with longer follow-up after implantation of an implantable cardioverter-defibrillator is increasing continuously. Defibrillation lead failure is a typical long-term complication. Therefore, the long-term reliability of implantable cardioverter-defibrillator leads has become an increasing concern. The aim of the present study was to assess the annual rate of transvenous defibrillation lead defects related to follow-up time after lead implantation.
Methods and results: A total of 990 consecutive patients who underwent first implantation of an implantable cardioverter-defibrillator between 1992 and May 2005 were analyzed. Median follow-up time was 934 days (interquartile range, 368 to 1870). Overall, 148 defibrillation leads (15%) failed during the follow-up. The estimated lead survival rates at 5 and 8 years after implantation were 85% and 60%, respectively. The annual failure rate increased progressively with time after implantation and reached 20% in 10-year-old leads (P<0.001). Lead defects affected newer as well as older models. Patients with lead defects were 3 years younger at implantation and more often female. Multiple lead implantation was associated with a trend to a higher rate of defibrillation lead defects (P=0.06). The major lead complications were insulation defects (56%), lead fractures (12%), loss of ventricular capture (11%), abnormal lead impedance (10%), and sensing failure (10%).
Conclusions: An increasing annual lead failure rate is noted primarily during long-term follow-up and reached 20% in 10-year-old leads. Patients with lead defects are younger and more often female.