Objectives: This study aims to compare automatic oscillometric blood pressure recordings with simultaneous direct intra-arterial blood pressure measurements in hyperacute stroke patients to test the accuracy of oscillometric readings.
Methods: A total of 51 first-ever stroke patients underwent simultaneous noninvasive automatic oscillometric and intra-arterial blood pressure monitoring within 3 h of ictus. Casual blood pressure was measured in both arms using a standard mercury sphygmomanometer on hospital admission. Patients who received antihypertensive medication during the blood pressure monitoring were excluded.
Results: The estimation of systolic blood pressure (SBP) using oscillometric recordings underestimated direct radial artery SBP by 9.7 mmHg (95% confidence interval: 6.5-13.0, P<0.001). In contrast, an upward bias of 5.6 mmHg (95% confidence interval: 3.5-7.7, P<0.001) was documented when noninvasive diastolic blood pressure (DBP) recordings were compared with intra-arterial DBP recordings. For SBP and DBP, the Pearson correlation coefficients between noninvasive and intra-arterial recordings were 0.854 and 0.832, respectively. When the study population was stratified according to SBP bands (group A: SBP<or=160 mmHg; group B: SBP>160 mmHg and SBP<or=180 mmHg, group C: SBP>180 mmHg), higher mean DeltaSBP (intra-arterial SBP-oscillometric SBP) levels were documented in group C (+19.8 mmHg, 95% confidence intervals: 12.2-27.4) when compared with groups B (+8.5 mmHg, 95% confidence intervals: 2.7-14.5; P=0.025) and A (+5.9 mmHg, 95% confidence intervals: 1.8-9.9; P=0.002).
Conclusion: Noninvasive automatic oscillometric BP measurements underestimate direct SBP recordings and overestimate direct DBP readings in acute stroke. The magnitude of the discrepancy between intra-arterial and oscillometric SBP recordings is even more prominent in patients with critically elevated SBP levels.