Distal left circumflex coronary artery flow reserve recorded by transthoracic Doppler echocardiography: a comparison with Doppler-wire

Cardiovasc Ultrasound. 2007 Jun 16:5:22. doi: 10.1186/1476-7120-5-22.

Abstract

Background: Coronary flow reserve (CFR) recording by means of transthoracic echocardiography (TTDE) in all the main distal coronary arteries is a challenge for advanced echocardiography. Validation studies of TTDE versus Doppler-wire (DW) recordings are available for Left Anterior Descending artery (LAD) and the Posterior Descending coronary artery (PD), but lacking for the more technically challenging Left Circumflex coronary artery (LCx).

Aim: To evaluate the reliability of TTDE in assessing CFR in LCx when compared to the intracoronary Doppler flow-wire gold standard.

Methods: we evaluated 5 patients (age = 60 +/- 9 years, 5 males) on LCx by TTDE and invasive CFR assessment. TTDE recording was performed using a low-frequency probe, with a four-chamber as a guiding 2D view. The 2 tests were performed on different days and in random order within 48 hours in a blind fashion. Vasodilator stimulus was adenosine, intravenously (140 gamma/kg/min x 3-6 min) for TTDE and intracoronary (40 gamma bolus) for DW recordings.

Results: CFR values on LCx ranged from 1.9 to 2.8 for DW, and from 2.0 to 3.0 for TTDE, with an overall correlation of R = 0,85 (p = 0,06); normal (CFR > 2.5) or abnormal (CFR < 2.5) value was concordantly identified by the 2 techniques in 4 out 5 cases (80%).

Conclusion: CFR of LCx artery can be obtained noninvasively with TTDE.

Publication types

  • Randomized Controlled Trial

MeSH terms

  • Aged
  • Coronary Angiography
  • Coronary Circulation*
  • Coronary Vessels / diagnostic imaging*
  • Coronary Vessels / physiology*
  • Coronary Vessels / physiopathology
  • Echocardiography / standards*
  • Humans
  • Male
  • Medical Records
  • Middle Aged
  • Myocardial Infarction / diagnostic imaging
  • Myocardial Infarction / physiopathology
  • Ultrasonography, Doppler / standards