Assessment of left atrial volumes at 1.5 Tesla and 3 Tesla using FLASH and SSFP cine imaging

J Cardiovasc Magn Reson. 2007;9(4):673-9. doi: 10.1080/10976640601138805.

Abstract

Purpose: To investigate left atrial volumes and function and their variability in healthy volunteers using steady state free precession (SSFP) and fast low angle shot (FLASH) sequences at both 1.5 and 3 T using both the short-axis and biplane area-length methods.

Materials and methods: Ten healthy volunteers underwent CMR at both 1.5 and 3 Tesla. The biplane area-length method utilized volumes from the horizontal and vertical long axis images.

Results: There were no significant differences between left atrial short-axis volumes or function between 1.5 and 3 T assessed using either FLASH or SSFP sequences. The biplane area-length method underestimated maximal left atrial volume using FLASH by 12 mL at 3 T (18%) and by 10 mL (14%) at 1.5 T (p = 0.003 and p = 0.05 respectively). Variability was larger for left atrial measurements using the biplane area-length method.

Conclusion: Field strength had no effect on left atrial volume and function assessment using either FLASH or SSFP. The use of the short-axis method for the acquisition of left atrial parameters is more reproducible than the biplane area-length for serial measurements.